Multi-Function Meters For Batch Control, Rate Indication and TotalisationP5000 Series
|
Model P5100 Shown Shown with FP7000 Series paddlewheel flow sensor (sold separately). Consult Sales Department for details. |
FREQUENCY INPUT TTL INPUT Speed and Protection Levels (jumper selectable): 7 MHz, 0-5 V; 100 kHz, 20 to +25 V; 3 kHz, 20 to +25 V Sensitivity (square-wave input): ±10 mV, 0 to 1 kHz; ±25 mV, 0 to 1 kHz; ±50 mV at 100 kHz Requires PGA1 NPN or PNP Open-Collector Sensor Excitation Output:12.4 V at 20 mA COMMON SPECIFICATIONS ALL INPUT TYPES Number of Inputs: 1 or 2 (2 inputs for frequency ratio and time interval only) Update Rate: 60 msec to 99.99 sec, field programmable OPERATION MODES Frequency/Tachometer Mode Frequency Range: 10-6 Hz to 7 MHz Accuracy at 25°C (square wave): ±0.0002% (+2 ppm) totalizer Display Offset (preset): 99,999 to 999,999 |
ANALOGUE INPUT Accuracy: better than 99.9% Non-Linearity: 0.05% FS Isolation: 350 Vdc between output and input Power: 230 Vac, 10-32 Vdc AC Frequency: 49-440 Hz Power Consumption, Typical: 3 W Battery Backup: user-supplied 6-12 Vdc, 60 mA to maintain operation, 400 mA with display Dimensions: 48 H x 96 W x 150 D mm (1.9" x 3.8" x 5.9"); 45 x 92 mm (1.8" x 3.6") cutout ON/OFF CONTROL & ALARM OUTPUTS Standard: three open-collector transistors, rated 150 mA sink, 30V Optional: two form C relays rated 8 A, 30 Vdc, or 240 Vac, resistive load (for rate alarm or batch control) Analogue Output Optional: isolated, scalable, internally powered and field selectable for 0-10 V, 4-20 mA, or 0-20 mA. Rangeable over 4 left most or right most digits, suitable for rate or total display. Min. Impedance for 10 V: 500 Max. External Impedance for 20 mA: 600 |
To Order (Specify Model Number) Prices Shown in Pounds | |||
---|---|---|---|
Model Number | Output Option | Input Option | Price |
P5100 P5101 P5102 P5103 P5104 |
None None None None None |
0 1 2 3 4 |
£285 368 350 385 399 |
P5110 P5111 P5112 P5113 P5114 |
BCD BCD BCD BCD BCD |
0 1 2 3 4 |
350 453 465 430 484 |
P5120 P5121 P5122 P5123 P5124 |
Relay Relay Relay Relay Relay |
0 1 2 3 4 |
355 430 440 405 460 |
P5130 P5131 P5132 P5133 P5134 |
Analogue Analogue Analogue Analogue Analogue |
0 1 2 3 4 |
415 495 499 465 519 |
Input Options:
'0' = Standard input, dual-channel input board, non-isolated TTL/5 V CMOS-level, frequency range of 10-6 to 7 MHz. Default calibration: 0 to 10 kHz = 0 to 10000.0.
'1' = Isolated single-channel input board with sensor excitation, protected for signals up to 260 Vac. (P6A1). Default calibration:
0 to 10 kHz = 0 to 10000.0.
'2' = Non-isolated dual-channel input board with sensor excitation. Protected for input signals up to 260 Vac RMS. (P6A2). Default calibration:
0 to 10 kHz = 0 to 10000.0
'3' = Non-isolated single-channel input board with sensor excitation. Protected for input signals up to 60 Vac RMS. (P6A3). Default calibration: 0 to 10 kHz = 0 to 10000.0
'4' = Isolated analogue-to-frequency single-channel input with 24 Vdc, 0-1 mA, 4-20 mA dc, 1-5 Vdc, and 0-10 Vdc (P6A5).
Highlights | Specifications | To Order | Product Information | Home |